<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//TaxonX//DTD Taxonomic Treatment Publishing DTD v0 20100105//EN" "https://foliamedica.bg/nlm/tax-treatment-NS0.dtd">
<article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:tp="http://www.plazi.org/taxpub" article-type="research-article" dtd-version="3.0" xml:lang="en">
  <front>
    <journal-meta>
      <journal-id journal-id-type="publisher-id">87</journal-id>
      <journal-id journal-id-type="index">urn:lsid:arphahub.com:pub:A116C711-4C18-5A38-8F1E-5E97753A8A64</journal-id>
      <journal-title-group>
        <journal-title xml:lang="en">Folia Medica</journal-title>
        <abbrev-journal-title xml:lang="en">FM</abbrev-journal-title>
      </journal-title-group>
      <issn pub-type="ppub">0204-8043</issn>
      <issn pub-type="epub">1314-2143</issn>
      <publisher>
        <publisher-name>Plovdiv Medical University</publisher-name>
      </publisher>
    </journal-meta>
    <article-meta>
      <article-id pub-id-type="doi">10.3897/folmed.67.e153885</article-id>
      <article-id pub-id-type="publisher-id">153885</article-id>
      <article-categories>
        <subj-group subj-group-type="heading">
          <subject>Research Article</subject>
        </subj-group>
        <subj-group subj-group-type="scientific_subject">
          <subject>Radiology &amp; Imaging</subject>
        </subj-group>
      </article-categories>
      <title-group>
        <article-title>Optimizing CT imaging: a cutting-edge validation of quality assurance across scanners</article-title>
      </title-group>
      <contrib-group content-type="authors">
        <contrib contrib-type="author" corresp="no">
          <name name-style="western">
            <surname>Baskar</surname>
            <given-names>Preethi</given-names>
          </name>
          <uri content-type="orcid">https://orcid.org/0009-0007-2034-7334</uri>
          <xref ref-type="aff" rid="A1">1</xref>
        </contrib>
        <contrib contrib-type="author" corresp="yes">
          <name name-style="western">
            <surname>Sivapadham</surname>
            <given-names>Sundaresan</given-names>
          </name>
          <email xlink:type="simple">preethib@srmist.edu.in</email>
          <xref ref-type="aff" rid="A2">2</xref>
        </contrib>
      </contrib-group>
      <aff id="A1">
        <label>1</label>
        <addr-line content-type="verbatim">Department of Radiodiagnosis, SRM Medical College Hospital &amp; Research Centre, SRM IST, SRM Nagar, Kattankulathur, Chengalpattu (Dt), Tamil Nadu – 603203</addr-line>
        <institution>Department of Radiodiagnosis, SRM Medical College Hospital &amp; Research Centre</institution>
        <addr-line content-type="city">Tamil Nadu</addr-line>
        <country>India</country>
      </aff>
      <aff id="A2">
        <label>2</label>
        <addr-line content-type="verbatim">Division of Medical Research, SRM Medical College Hospital &amp; Research Centre, SRM IST, SRM Nagar, Kattankulathur, Chengalpattu (Dt), Tamil Nadu – 603203</addr-line>
        <institution>Division of Medical Research, SRM Medical College Hospital &amp; Research Centre</institution>
        <addr-line content-type="city">Tamil Nadu</addr-line>
        <country>India</country>
      </aff>
      <author-notes>
        <fn fn-type="corresp">
          <p>Corresponding author: Sundaresan Sivapatham, SRM University, Chengalpattu, Tamil Nadu, India; Email: <email xlink:type="simple">preethib@srmist.edu.in</email></p>
        </fn>
      </author-notes>
      <pub-date pub-type="collection">
        <year>2025</year>
      </pub-date>
      <pub-date pub-type="epub">
        <day>22</day>
        <month>10</month>
        <year>2025</year>
      </pub-date>
      <volume>67</volume>
      <issue>5</issue>
      <elocation-id>e153885</elocation-id>
      <uri content-type="arpha" xlink:href="http://openbiodiv.net/34256906-D1F0-5270-9F45-95E001375950">34256906-D1F0-5270-9F45-95E001375950</uri>
      <history>
        <date date-type="received">
          <day>25</day>
          <month>03</month>
          <year>2025</year>
        </date>
        <date date-type="accepted">
          <day>01</day>
          <month>06</month>
          <year>2025</year>
        </date>
      </history>
      <permissions>
        <copyright-statement>Preethi Baskar, Sundaresan Sivapadham</copyright-statement>
        <license license-type="creative-commons-attribution" xlink:href="http://creativecommons.org/licenses/by/4.0/" xlink:type="simple">
          <license-p>This is an open access article distributed under the terms of the Creative Commons Attribution License (CC BY 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.</license-p>
        </license>
      </permissions>
      <abstract>
        <p>
          <bold>Abstract</bold>
        </p>
        <p><bold>Introduction</bold>: A comprehensive test on a dual- and multi-slice <abbrev xlink:title="Computed tomography" id="ABBRID0END">CT</abbrev> scanner was carried out using in-house <abbrev xlink:title="Computed tomography" id="ABBRID0ERD">CT</abbrev> scanning with a <abbrev xlink:title="computed tomography dose index" id="ABBRID0EVD">CTDI</abbrev> phantom/test tool following the American Association of Physicists in Medicine’s recommended methods to validate quality assurance of simulation <abbrev xlink:title="Computed tomography" id="ABBRID0EZD">CT</abbrev> and diagnostic <abbrev xlink:title="Computed tomography" id="ABBRID0E4D">CT</abbrev>.</p>
        <p><bold>Aim</bold>: To carry out the mechanical and radiation dose verification <abbrev xlink:title="quality assurance" id="ABBRID0EFE">QA</abbrev> tests for the Siemens Emotion Duo <abbrev xlink:title="Computed tomography" id="ABBRID0EJE">CT</abbrev> simulator and GE <abbrev xlink:title="Computed tomography" id="ABBRID0ENE">CT</abbrev> as per the <abbrev xlink:title="Atomic Energy Regulatory Board" id="ABBRID0ERE">AERB</abbrev><abbrev xlink:title="quality assurance" id="ABBRID0EVE">QA</abbrev> protocol.</p>
        <p><bold>Materials and methods</bold>: The Siemens Emotion Duo is a twin-slice <abbrev xlink:title="Computed tomography" id="ABBRID0E4E">CT</abbrev> scanner that uses UFC dual-detector technology to provide sub-second scanning for a wide range of clinical applications. The precision-killed <abbrev xlink:title="polymethyl methacrylate" id="ABBRID0EBF">PMMA</abbrev> (polymethyl methacrylate) nested <abbrev xlink:title="computed tomography dose index" id="ABBRID0EFF">CTDI</abbrev> phantom has a 32 cm diameter body phantom and a 16 cm diameter head phantom. For the positioning of the <abbrev xlink:title="polymethyl methacrylate" id="ABBRID0EJF">PMMA</abbrev> for the <abbrev xlink:title="Computed tomography" id="ABBRID0ENF">CT</abbrev> ion chamber, the phantom is built with a central hole and either peripheral hole (four in the head phantom).</p>
        <p><bold>Results</bold>: Comprehensive <abbrev xlink:title="quality assurance" id="ABBRID0EVF">QA</abbrev> and validation of results are necessary to obtain baseline data for <abbrev xlink:title="Computed tomography" id="ABBRID0EZF">CT</abbrev> virtual simulation and diagnostics. The results of <abbrev xlink:title="quality assurance" id="ABBRID0E4F">QA</abbrev> data for <abbrev xlink:title="Computed tomography" id="ABBRID0EBG">CT</abbrev> equipment are compared and found to be within acceptable tolerance limits.</p>
        <p><bold>Conclusion</bold>: The results are unaffected by differences in collimations, reconstruction algorithms, or spectral energy because the data were collected over time on various <abbrev xlink:title="Computed tomography" id="ABBRID0EJG">CT</abbrev> scanners using the Piranha dosimetry.</p>
      </abstract>
      <kwd-group>
        <label>Keywords</label>
        <kwd>CT simulator</kwd>
        <kwd>Siemens Somatom Emotion DUO</kwd>
        <kwd>GE CT CTDI phantom</kwd>
        <kwd>0.1 cc ion chamber</kwd>
        <kwd>Piranha 657</kwd>
      </kwd-group>
    </article-meta>
    <notes>
      <sec sec-type="Citation" id="SECID0EVG">
        <title>Citation</title>
        <p>Baskar P, Sivapadham S. Optimizing <abbrev xlink:title="Computed tomography" id="ABBRID0E2G">CT</abbrev> imaging: a cutting-edge validation of quality assurance across scanners. Folia Med (Plovdiv) 2025;67(5):е153885. doi: <ext-link xlink:type="simple" ext-link-type="doi" xlink:href="10.3897/folmed.67.e153885">10.3897/folmed.67.e153885</ext-link>.</p>
      </sec>
    </notes>
  </front>
  <body>
    <sec sec-type="Introduction" id="SECID0EFH">
      <title>Introduction</title>
      <p>A quality assurance program has the potential to contribute significantly to the production of images of consistently high quality with minimum patient exposure. This is achieved by aiding in the detection of changes in the image-producing variable and adversely affecting the quality of the image. Poor image quality can impede the accessibility of diagnostic information for medical professionals, thereby elevating the likelihood of an erroneous diagnosis. If a diagnosis is made on the basis of a poor image, the radiation exposure the patient has received will have been entirely superfluous. In the event that a radiography is repeated, the patient’s exposure is naturally doubled. Data from the surveys suggest that some patients may be subjected to unnecessary radiation exposure on a regular basis. The extent of radiation exposure for a given patient can vary significantly between facilities, with some facilities reporting exposure levels that are 50 times higher than others. As with all imaging systems, the <abbrev xlink:title="Computed tomography" id="ABBRID0ELH">CT</abbrev> system necessitates the execution of quality assurance (<abbrev xlink:title="quality assurance" id="ABBRID0EPH">QA</abbrev>) tests on a daily, monthly, and yearly basis.<sup>[<xref ref-type="bibr" rid="B1">1</xref>]</sup> Computed tomography (<abbrev xlink:title="Computed tomography" id="ABBRID0E1H">CT</abbrev>) scanning is a diagnostic imaging procedure that utilizes X-rays and advanced computerized technology to generate cross-sectional images of the body from multiple angles. The Piranha 657 model has been designed for utilization in a variety of applications, including <abbrev xlink:title="Computed tomography" id="ABBRID0E5H">CT</abbrev>, fluoroscopy, mammography, and radiography. The device is capable of measuring <abbrev xlink:title="half-value layer" id="ABBRID0EDAAC">HVL</abbrev>, total filtration, time, dose/dose rate, and kVp. The same model dosimetry has been applied to diagnostic and simulation <abbrev xlink:title="Computed tomography" id="ABBRID0EHAAC">CT</abbrev> in four centers for quality assurance <bold>(Table <xref ref-type="table" rid="T1">1</xref>)</bold>.</p>
      <table-wrap id="T1" position="float" orientation="portrait">
        <label>Table 1.</label>
        <caption>
          <p>Details of computer tomography scanners</p>
        </caption>
        <table id="TID0E4EAE" rules="all">
          <tbody>
            <tr>
              <td rowspan="1" colspan="1">
                <bold>Make</bold>
              </td>
              <td rowspan="1" colspan="1">
                <bold>Slice</bold>
              </td>
              <td rowspan="1" colspan="1">
                <bold>Model</bold>
              </td>
              <td rowspan="1" colspan="1">
                <bold>Scan (t) (s)</bold>
              </td>
              <td rowspan="1" colspan="1">
                <bold>Hospitals</bold>
              </td>
              <td rowspan="1" colspan="1">
                <bold>Machine denoted as</bold>
              </td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">Siemens</td>
              <td rowspan="1" colspan="1">16</td>
              <td rowspan="1" colspan="1">Somatom Emotion</td>
              <td rowspan="1" colspan="1">0.6–1.5</td>
              <td rowspan="1" colspan="1">GAACH &amp; RC Kancheepuram Tamilnadu, India</td>
              <td rowspan="1" colspan="1">A</td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">Siemens</td>
              <td rowspan="1" colspan="1">32</td>
              <td rowspan="1" colspan="1">Somatom Spirit and 21109</td>
              <td rowspan="1" colspan="1">0.8</td>
              <td rowspan="1" colspan="1">SRM MCH &amp; RC Kattangulathur Chennai, India</td>
              <td rowspan="1" colspan="1">B</td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">GE</td>
              <td rowspan="1" colspan="1">32</td>
              <td rowspan="1" colspan="1">REV ACTs Expert Edition</td>
              <td rowspan="1" colspan="1">0.98</td>
              <td rowspan="1" colspan="1">SRM Global Kattangulathur Chennai, India</td>
              <td rowspan="1" colspan="1">E</td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">GE</td>
              <td rowspan="1" colspan="1">128</td>
              <td rowspan="1" colspan="1">Optima CT660</td>
              <td rowspan="1" colspan="1">0.675</td>
              <td rowspan="1" colspan="1">SRM MCH &amp; RC Chengalpattu, India</td>
              <td rowspan="1" colspan="1">D</td>
            </tr>
          </tbody>
        </table>
      </table-wrap>
      <p>Several key scan parameters strongly influence <abbrev xlink:title="Computed tomography" id="ABBRID0EUEAC">CT</abbrev> image quality and the possibility of artifacts. Noise levels and tissue contrast are influenced by tube current (mA) and voltage (kVp), as slice depth and pitch affect spatial resolution and scan time, respectively. Higher mA and thicker slices reduce noise while increasing irradiance dose, but lower mA and thinner slices can lead to grainy images and motion distortion.<sup>[<xref ref-type="bibr" rid="B2">2</xref>]</sup> Incorrect choice of reconstruction algorithms and field of view can lead to loss of detail or aliasing. Furthermore, patient motion and incorrect positioning can lead to blurring and banding artifacts. As a result, careful adjustment of these parameters is necessary to ensure a high-quality image while reducing artifacts.</p>
      <p>The American Association of Physicists in Medicine (<abbrev xlink:title="American Association of Physicists in Medicine" id="ABBRID0EBFAC">AAPM</abbrev>) and Atomic Energy Regulatory Board (<abbrev xlink:title="Atomic Energy Regulatory Board" id="ABBRID0EFFAC">AERB</abbrev>) protocols were applied to examine four <abbrev xlink:title="Computed tomography" id="ABBRID0EJFAC">CT</abbrev> units from various models. These protocols included testing for accelerating voltage (kVp), mAs accuracy and its corresponding dose, half-value layer (<abbrev xlink:title="half-value layer" id="ABBRID0ENFAC">HVL</abbrev>) measurement, machine output using a Piranha multimeter, and <abbrev xlink:title="computed tomography dose index" id="ABBRID0ERFAC">CTDI</abbrev> dose measurement using both head and body <abbrev xlink:title="computed tomography dose index" id="ABBRID0EVFAC">CTDI</abbrev> phantoms with 0.1 cc ion chambers.<sup>[<xref ref-type="bibr" rid="B3">3</xref>]</sup></p>
      <p>The <abbrev xlink:title="Computed tomography" id="ABBRID0EBGAC">CT</abbrev> scanner we used in the study employs the UFC dual detector technology to provide sub-second scanning for a wide range of clinical applications.<sup>[<xref ref-type="bibr" rid="B4">4</xref>]</sup> For excellent image reformatting, the Siemens Emotion Duo, a dual-slice <abbrev xlink:title="Computed tomography" id="ABBRID0EMGAC">CT</abbrev> scanner with UFC detector technology, offers anatomical volumes with thinner slices. In comparison to the single-slice scanners, it uses higher mA to enhance image quality and reduce X-ray tube load.<sup>[<xref ref-type="bibr" rid="B5">5</xref>]</sup> This spiral <abbrev xlink:title="Computed tomography" id="ABBRID0EXGAC">CT</abbrev> scanner has a 3.5 MHU tube, an 80-cm gantry aperture, and TFT rotation times of 0.8, 1, and 1.5 seconds. Slice thickness ranges from 1 to 8 mm, and gantry tilt ranges from −15° to +15°. Diagnostic <abbrev xlink:title="Computed tomography" id="ABBRID0E2GAC">CT</abbrev> scanners, except for gantry aperture and couch design, share similar specifications. The diagnostic <abbrev xlink:title="Computed tomography" id="ABBRID0E6GAC">CT</abbrev> gantry has a 68.58 cm diameter and a curved couch. The HPIPS <abbrev xlink:title="Computed tomography" id="ABBRID0EDHAC">CT</abbrev> scanner China has been using a sophisticated GE <abbrev xlink:title="Computed tomography" id="ABBRID0EHHAC">CT</abbrev> scanner device for picture collection. The GE <abbrev xlink:title="Computed tomography" id="ABBRID0ELHAC">CT</abbrev> scanner model has been analyzed using a convolutional neural network, which enhances the <abbrev xlink:title="Computed tomography" id="ABBRID0EPHAC">CT</abbrev> scanner’s quality. It performs fast, low-dose scans without surrendering speed, power, or field of view.</p>
    </sec>
    <sec sec-type="Aim" id="SECID0ETHAC">
      <title>Aim</title>
      <p>The aim of the study was to assess whether the Piranha dosimeter approach is a reliable, sensitive, and acceptable <abbrev xlink:title="quality assurance" id="ABBRID0EZHAC">QA</abbrev> test for AEC systems from four leading <abbrev xlink:title="Computed tomography" id="ABBRID0E4HAC">CT</abbrev> manufacturers. This test can be done in each different machine (i.e., A, B, C, D, and E).</p>
      <sec sec-type="The precision PMMA (polymethyl methacrylate) nested CTDI Phantom" id="SECID0EBIAC">
        <title>The precision PMMA (polymethyl methacrylate) nested CTDI Phantom</title>
        <p>The <abbrev xlink:title="polymethyl methacrylate" id="ABBRID0EQIAC">PMMA</abbrev> nested <abbrev xlink:title="computed tomography dose index" id="ABBRID0EUIAC">CTDI</abbrev> phantom, widely available in radiology departments, includes a central hole and four peripheral holes in the head phantom for <abbrev xlink:title="Computed tomography" id="ABBRID0EYIAC">CT</abbrev> ion chamber positioning. For <abbrev xlink:title="computed tomography dose index" id="ABBRID0E3IAC">CTDI</abbrev> measurements, the DTC10 is a tough, cylindrical, pencil-shaped air ionization chamber <bold>(Fig. <xref ref-type="fig" rid="F1">1</xref>)</bold>. The <abbrev xlink:title="Computed tomography" id="ABBRID0EHJAC">CT</abbrev> chamber is used to measure and track the output amount of exposure from <abbrev xlink:title="Computed tomography" id="ABBRID0ELJAC">CT</abbrev> scanners in a phantom or in an open environment. The IEC 61223-2-6 standard can be used to measure the <abbrev xlink:title="Computed tomography" id="ABBRID0EPJAC">CT</abbrev> dose index (<abbrev xlink:title="computed tomography dose index" id="ABBRID0ETJAC">CTDI</abbrev>). DTC10 can be connected to the Piranha to measure <abbrev xlink:title="computed tomography dose index" id="ABBRID0EXJAC">CTDI</abbrev> using a chamber adapter.<sup>[<xref ref-type="bibr" rid="B6">6</xref>]</sup> It is possible to link the DTC10 directly to the barracuda (using an electrometer module that supports bias measurement of the <abbrev xlink:title="computed tomography dose index" id="ABBRID0ECKAC">CTDI</abbrev>). The specifications of the 0.1 cc ionization chamber include a dose range of −0.8 mGy cm/s to 8 mGy cm/s, a length rate of Gy cm/s, an inaccuracy of ±6% or ±0.08 mGy cm, a 2 m low-noise triaxial cable, an active volume of 4.9 cm, an active length of 100 mm, an inner diameter of 8.0 cm, and typical leakage of ±4 fA.<sup>[<xref ref-type="bibr" rid="B7">7</xref>]</sup> Total filtration can be verified over a 100-meter Bluetooth range. <abbrev xlink:title="quality assurance" id="ABBRID0ENKAC">QA</abbrev> experiments for the Siemens Emotion Duo and GE <abbrev xlink:title="Computed tomography" id="ABBRID0ERKAC">CT</abbrev> scanners followed the TG-66 protocol for both diagnostic and simulation <abbrev xlink:title="Computed tomography" id="ABBRID0EVKAC">CT</abbrev>.</p>
        <fig id="F1" position="float" orientation="portrait">
          <object-id content-type="arpha">96031BD2-9616-5BC4-8C75-61E096B947D9</object-id>
          <label>Figure 1.</label>
          <caption>
            <p>Centre and peripheral <abbrev xlink:title="computed tomography dose index" id="ABBRID0EBLAC">CTDI</abbrev> calculated phantom setup.</p>
          </caption>
          <graphic xlink:href="foliamedica-67-5-e153885-g001.jpg" position="float" orientation="portrait" xlink:type="simple" id="oo_1446610.jpg">
            <uri content-type="original_file">https://binary.pensoft.net/fig/1446610</uri>
          </graphic>
        </fig>
      </sec>
      <sec sec-type="Dosimetry – CTDI in air" id="SECID0EKLAC">
        <title>Dosimetry – CTDI in air</title>
        <p>A 10 cm pencil-shaped ionization chamber, positioned near the scan isocenter, quantifies the computed tomography dose index (<abbrev xlink:title="computed tomography dose index" id="ABBRID0EVLAC">CTDI</abbrev>). A laboratory testing stand and a clamp support the ion chamber. The <abbrev xlink:title="computed tomography dose index" id="ABBRID0EZLAC">CTDI</abbrev> is expressed as<sup>[<xref ref-type="bibr" rid="B8">8</xref>]</sup>:</p>
        <p>
          <mml:math id="M1">
            <mml:mi>CTDI</mml:mi>
            <mml:mo>=</mml:mo>
            <mml:mn>1</mml:mn>
            <mml:mo>/</mml:mo>
            <mml:mi mathvariant="normal">T</mml:mi>
            <mml:msubsup>
              <mml:mo>∫</mml:mo>
              <mml:mrow>
                <mml:mo>−</mml:mo>
                <mml:mi mathvariant="normal">∞</mml:mi>
              </mml:mrow>
              <mml:mrow>
                <mml:mo>+</mml:mo>
                <mml:mi mathvariant="normal">∞</mml:mi>
              </mml:mrow>
            </mml:msubsup>
            <mml:mi mathvariant="normal">D</mml:mi>
            <mml:mo>(</mml:mo>
            <mml:mi mathvariant="normal">Z</mml:mi>
            <mml:mo>)</mml:mo>
            <mml:mi>dz</mml:mi>
          </mml:math>
        </p>
        <p>where T is the nominal slice thickness, and D(z) is the dose profile along the tube’s rotation axis. In helical mode, the <abbrev xlink:title="computed tomography dose index" id="ABBRID0ECNAC">CTDI</abbrev> is adjusted for overlapping slices by dividing by the pitch factor to calculate the multiple scan average dose (<abbrev xlink:title="multiple scan average dose" id="ABBRID0EGNAC">MSAD</abbrev>):</p>
        <p><abbrev xlink:title="multiple scan average dose" id="ABBRID0EMNAC">MSAD</abbrev> = <abbrev xlink:title="computed tomography dose index" id="ABBRID0EQNAC">CTDI</abbrev> /P</p>
        <p>where pitch (P) is defined as r = Table Feed Per Rotation divided by Nominal Slice Thickness. The <abbrev xlink:title="computed tomography dose index" id="ABBRID0EWNAC">CTDI</abbrev> phantom is centered and aligned at the scan isocenter on the patient table. The ion chamber is inserted to measure the central <abbrev xlink:title="computed tomography dose index" id="ABBRID0E1NAC">CTDI</abbrev>, and peripheral <abbrev xlink:title="computed tomography dose index" id="ABBRID0E5NAC">CTDI</abbrev> is measured at least four times around the phantom to calculate an average.</p>
      </sec>
    </sec>
    <sec sec-type="Results" id="SECID0ECOAC">
      <title>Results</title>
      <sec sec-type="Laser QA procedure" id="SECID0EGOAC">
        <title>Laser QA procedure</title>
        <p>The laser <abbrev xlink:title="quality assurance" id="ABBRID0EROAC">QA</abbrev> device is attached to the simulator tabletop, and the couch is adjusted until the side holes in the pegs align with the horizontal and vertical wall lasers.<sup>[<xref ref-type="bibr" rid="B8">8</xref>]</sup> Then both left and right wall lasers are verified to align with the holes within a tolerance of ±2 mm.</p>
      </sec>
      <sec sec-type="Scanner table test" id="SECID0E3OAC">
        <title>Scanner table test</title>
        <p>The laser <abbrev xlink:title="quality assurance" id="ABBRID0ECPAC">QA</abbrev> tool is placed as close as possible to the head of the tabletop (gantry side) and aligned with the gantry laser to capture a single image. The device is then moved as close as possible to the table’s base and realigned with the gantry laser once more to capture a single image.<sup>[<xref ref-type="bibr" rid="B9">9</xref>]</sup> Two photos should be taken from the same position for the laser <abbrev xlink:title="quality assurance" id="ABBRID0ENPAC">QA</abbrev> device. The center hole in the <abbrev xlink:title="quality assurance" id="ABBRID0ERPAC">QA</abbrev> device is located using the scanner cursor tool on both images. The location of the hole should be within 2 mm of the location shown in both photos.</p>
      </sec>
      <sec sec-type="Gantry tilt" id="SECID0EVPAC">
        <title>Gantry tilt</title>
        <p>A square plastic sheet with water-equivalent properties is affixed to the pre-packed X-ray film and positioned with the sagittal gantry laser intersecting the film’s vertical axis. A single scan is acquired at a 0° gantry angle with the narrowest slice thickness. The gantry is then tilted in both directions, and single scans are acquired at each position, near the limit of gantry movement.<sup>[<xref ref-type="bibr" rid="B10">10</xref>]</sup> The angle between the vertical and tilted exposures, measured with a protractor, should agree within ±1° of the digitally indicated gantry angle.<sup>[<xref ref-type="bibr" rid="B11">11</xref>]</sup> Resulting images show anterior-posterior (<abbrev xlink:title="anterior-posterior" id="ABBRID0EKAAE">AP</abbrev>) and lateral (<abbrev xlink:title="lateral" id="ABBRID0EOAAE">LAT</abbrev>) diameters. <bold>Fig. <xref ref-type="fig" rid="F2">2</xref></bold> shows gantry angles that vary from 0° to 15°. At 0°, the <abbrev xlink:title="anterior-posterior" id="ABBRID0EYAAE">AP</abbrev> and <abbrev xlink:title="lateral" id="ABBRID0E3AAE">LAT</abbrev> diameters are equal. Gantry angle results for different machines (0.2°, 0.4°, 0.5°) are listed in <bold>Table <xref ref-type="table" rid="T2">2</xref></bold>.</p>
        <table-wrap id="T2" position="float" orientation="portrait">
          <label>Table 2.</label>
          <caption>
            <p>A comparison of different <abbrev xlink:title="Computed tomography" id="ABBRID0EPBAE">CT</abbrev> dose quality and quantity factors (A, B, C, and D)</p>
          </caption>
          <table id="TID0E5LAE" rules="all">
            <tbody>
              <tr>
                <td rowspan="1" colspan="1">Type of machine denoted as</td>
                <td rowspan="1" colspan="1">Accuracy of gantry tilt (degree)</td>
                <td rowspan="1" colspan="1">Accuracy of slice thickness  (mm)</td>
                <td rowspan="1" colspan="1">Accuracy of operation potential (kV)</td>
                <td rowspan="1" colspan="1">Timer accuracy  (s)</td>
                <td rowspan="1" colspan="1">Reproducibility of output (CoV)</td>
                <td rowspan="1" colspan="1">Low contrast resolution (lp/mm)</td>
                <td rowspan="1" colspan="1">High contrast resolution (lp/cm)</td>
                <td rowspan="1" colspan="1"><abbrev xlink:title="computed tomography dose index" id="ABBRID0EIDAE">CTDI</abbrev>  Centre peripheral (mGy)</td>
                <td rowspan="1" colspan="1"><abbrev xlink:title="Computed tomography" id="ABBRID0EUDAE">CT</abbrev>  Number accuracy (5HU)</td>
                <td rowspan="1" colspan="1">Total filtration (mmAl)</td>
                <td rowspan="1" colspan="1">Radiation leakage)  (mR in hour)</td>
              </tr>
              <tr>
                <td rowspan="7" colspan="1">Siemens 32 (A)</td>
                <td rowspan="1" colspan="1">+15°</td>
                <td rowspan="1" colspan="1">0.625</td>
                <td rowspan="1" colspan="1">80</td>
                <td rowspan="1" colspan="1">1.00</td>
                <td rowspan="1" colspan="1">80</td>
                <td rowspan="4" colspan="6">120 kV, 100 mA</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">−15°</td>
                <td rowspan="1" colspan="1">1.000</td>
                <td rowspan="1" colspan="1">110</td>
                <td rowspan="1" colspan="1">1.50</td>
                <td rowspan="1" colspan="1">110</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">4.000</td>
                <td rowspan="1" colspan="1">130</td>
                <td rowspan="1" colspan="1">2.00</td>
                <td rowspan="1" colspan="1">120</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">5.000</td>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1"/>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">0.2°</td>
                <td rowspan="1" colspan="1">1.12</td>
                <td rowspan="1" colspan="1">78.33</td>
                <td rowspan="1" colspan="1">1.04</td>
                <td rowspan="1" colspan="1">0.0015</td>
                <td rowspan="3" colspan="1">4.00</td>
                <td rowspan="3" colspan="1">6.02</td>
                <td rowspan="1" colspan="1">Head 15.54</td>
                <td rowspan="1" colspan="1">+6</td>
                <td rowspan="3" colspan="1">4.86</td>
                <td rowspan="3" colspan="1">0.080</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">3.80</td>
                <td rowspan="1" colspan="1">109.6</td>
                <td rowspan="1" colspan="1">1.52</td>
                <td rowspan="1" colspan="1">0.0016</td>
                <td rowspan="1" colspan="1">Body 7.26</td>
                <td rowspan="1" colspan="1">−2</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">4.90</td>
                <td rowspan="1" colspan="1">128.3</td>
                <td rowspan="1" colspan="1">2.02</td>
                <td rowspan="1" colspan="1">0.0010</td>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1"/>
              </tr>
              <tr>
                <td rowspan="6" colspan="1">Siemens 16 (B)</td>
                <td rowspan="1" colspan="1">+15°</td>
                <td rowspan="1" colspan="1">02.0</td>
                <td rowspan="1" colspan="1">80</td>
                <td rowspan="1" colspan="1">1.00</td>
                <td rowspan="1" colspan="1">80</td>
                <td rowspan="3" colspan="6">120 kV, 100 mA</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">−15°</td>
                <td rowspan="1" colspan="1">05.0</td>
                <td rowspan="1" colspan="1">100</td>
                <td rowspan="1" colspan="1">1.50</td>
                <td rowspan="1" colspan="1">100</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">10.0</td>
                <td rowspan="1" colspan="1">120</td>
                <td rowspan="1" colspan="1">2.00</td>
                <td rowspan="1" colspan="1">120</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">0.4°</td>
                <td rowspan="1" colspan="1">02.47</td>
                <td rowspan="1" colspan="1">80.51</td>
                <td rowspan="1" colspan="1">0.30</td>
                <td rowspan="1" colspan="1">0.0003</td>
                <td rowspan="3" colspan="1">3.00</td>
                <td rowspan="3" colspan="1">6.20</td>
                <td rowspan="1" colspan="1">Head 14.94</td>
                <td rowspan="1" colspan="1">+7</td>
                <td rowspan="3" colspan="1">7.70</td>
                <td rowspan="3" colspan="1">0.48</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">05.05</td>
                <td rowspan="1" colspan="1">99.41</td>
                <td rowspan="1" colspan="1">0.30</td>
                <td rowspan="1" colspan="1">0.0021</td>
                <td rowspan="1" colspan="1">Body 8.55</td>
                <td rowspan="1" colspan="1">−3</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">10.02</td>
                <td rowspan="1" colspan="1">119.60</td>
                <td rowspan="1" colspan="1">0.30</td>
                <td rowspan="1" colspan="1">0.0018</td>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1"/>
              </tr>
              <tr>
                <td rowspan="6" colspan="1">GE 32 (C)</td>
                <td rowspan="1" colspan="1">+15°</td>
                <td rowspan="1" colspan="1">1.00</td>
                <td rowspan="1" colspan="1">80</td>
                <td rowspan="1" colspan="1">1.50</td>
                <td rowspan="1" colspan="1">80</td>
                <td rowspan="3" colspan="6">120 kV, 100 mA</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">−15°</td>
                <td rowspan="1" colspan="1">3.00</td>
                <td rowspan="1" colspan="1">100</td>
                <td rowspan="1" colspan="1">1.00</td>
                <td rowspan="1" colspan="1">100</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">5.00</td>
                <td rowspan="1" colspan="1">120</td>
                <td rowspan="1" colspan="1">0.50</td>
                <td rowspan="1" colspan="1">120</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">0.5°</td>
                <td rowspan="1" colspan="1">1.30</td>
                <td rowspan="1" colspan="1">79.78</td>
                <td rowspan="1" colspan="1">0.4%</td>
                <td rowspan="1" colspan="1">0.0023</td>
                <td rowspan="3" colspan="1">3.00</td>
                <td rowspan="3" colspan="1">6.50</td>
                <td rowspan="1" colspan="1">Body 5.29</td>
                <td rowspan="1" colspan="1">+8</td>
                <td rowspan="3" colspan="1">6.90</td>
                <td rowspan="3" colspan="1">0.006</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">3.10</td>
                <td rowspan="1" colspan="1">98.34</td>
                <td rowspan="1" colspan="1">0.4%</td>
                <td rowspan="1" colspan="1">0.0020</td>
                <td rowspan="1" colspan="1">Head 15.456</td>
                <td rowspan="1" colspan="1">−1</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">5.20</td>
                <td rowspan="1" colspan="1">119.7</td>
                <td rowspan="1" colspan="1">1.3%</td>
                <td rowspan="1" colspan="1">0.0010</td>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1"/>
              </tr>
              <tr>
                <td rowspan="6" colspan="1">GE 128 (D)</td>
                <td rowspan="1" colspan="1">+15°</td>
                <td rowspan="1" colspan="1">2.00</td>
                <td rowspan="1" colspan="1">80</td>
                <td rowspan="1" colspan="1">1.0</td>
                <td rowspan="1" colspan="1">80</td>
                <td rowspan="3" colspan="6">120 kV, 100 mA</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">−15</td>
                <td rowspan="1" colspan="1">3.00</td>
                <td rowspan="1" colspan="1">130</td>
                <td rowspan="1" colspan="1">1.5</td>
                <td rowspan="1" colspan="1">120</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">5.00</td>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">2.0</td>
                <td rowspan="1" colspan="1">130</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">0.3°</td>
                <td rowspan="1" colspan="1">2.08</td>
                <td rowspan="1" colspan="1">79.87</td>
                <td rowspan="1" colspan="1">0.35%</td>
                <td rowspan="1" colspan="1">0.005</td>
                <td rowspan="3" colspan="1">5.00</td>
                <td rowspan="3" colspan="1">8.40</td>
                <td rowspan="1" colspan="1">Head 25.48</td>
                <td rowspan="1" colspan="1">+6</td>
                <td rowspan="3" colspan="1">8.50</td>
                <td rowspan="3" colspan="1">0.023</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">2.96</td>
                <td rowspan="1" colspan="1">131.7</td>
                <td rowspan="1" colspan="1">0.35%</td>
                <td rowspan="1" colspan="1">0.002</td>
                <td rowspan="1" colspan="1">Body 13.71</td>
                <td rowspan="1" colspan="1">−5</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">4.44</td>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1">0.35%</td>
                <td rowspan="1" colspan="1">0.001</td>
                <td rowspan="1" colspan="1"/>
                <td rowspan="1" colspan="1"/>
              </tr>
            </tbody>
          </table>
        </table-wrap>
        <fig id="F2" position="float" orientation="portrait">
          <object-id content-type="arpha">34BF2A9B-1E12-5B00-A61D-F1B41E5B2B68</object-id>
          <label>Figure 2.</label>
          <caption>
            <p>Gantry tilt test for A, B, C, and D.</p>
          </caption>
          <graphic xlink:href="foliamedica-67-5-e153885-g002.jpg" position="float" orientation="portrait" xlink:type="simple" id="oo_1446611.jpg">
            <uri content-type="original_file">https://binary.pensoft.net/fig/1446611</uri>
          </graphic>
        </fig>
      </sec>
      <sec sec-type="Low contrast resolution" id="SECID0EUVAE">
        <title>Low contrast resolution</title>
        <p>The smallest discernible item at a specific contrast level for a given dose is often referred to as low contrast resolution (or low contrast detectability) (<abbrev xlink:title="low contrast resolution" id="ABBRID0E1VAE">LCR</abbrev>). The measurement is imperative to confirm during acceptance testing since it directly corresponds to imaging performance. <abbrev xlink:title="low contrast resolution" id="ABBRID0E5VAE">LCR</abbrev> is assessed by imaging a low-contrast insert in at least 20 images and reviewed by at least three observers under ideal conditions to generate an average.<sup>[<xref ref-type="bibr" rid="B12">12</xref>]</sup> The <abbrev xlink:title="low contrast resolution" id="ABBRID0EJWAE">LCR</abbrev> assessment by human observers demonstrated an overall increase in the mean reader score (better <abbrev xlink:title="low contrast resolution" id="ABBRID0ENWAE">LCR</abbrev>) with increasing dose. However, in this study, the same dose value was set as 100 mA and 120 kV. For vendors A, B, C, and D <bold>(Fig. <xref ref-type="fig" rid="F3">3</xref>)</bold>, the Picture Postcard Resolutions (<abbrev xlink:title="Picture Postcard Resolutions" id="ABBRID0EYWAE">PPR</abbrev>) for human observer assessment of the images’ <abbrev xlink:title="low contrast resolution" id="ABBRID0E3WAE">LCR</abbrev> were 10% (4.0), 8.3% (3.2), 8% (3), and 12% (5), respectively. Vendor D exhibited suboptimal contrast, yet its performance remained within the acceptable range of &lt;15%, as defined by the established criteria.</p>
        <fig id="F3" position="float" orientation="portrait">
          <object-id content-type="arpha">658BFBE1-D460-5476-9CE0-FF1C8F338082</object-id>
          <label>Figure 3.</label>
          <caption>
            <p>Low and high contrast test for A, B, C, and D.</p>
          </caption>
          <graphic xlink:href="foliamedica-67-5-e153885-g003.jpg" position="float" orientation="portrait" xlink:type="simple" id="oo_1446612.jpg">
            <uri content-type="original_file">https://binary.pensoft.net/fig/1446612</uri>
          </graphic>
        </fig>
        <p>The following report presents a summary of the mechanical safety performance test results for Somatom Siemens Medical <abbrev xlink:title="Computed tomography" id="ABBRID0EOXAE">CT</abbrev> and GE scanners.<sup>[<xref ref-type="bibr" rid="B13">13</xref>]</sup> The measured slice thickness accuracy of the <abbrev xlink:title="Computed tomography" id="ABBRID0EZXAE">CT</abbrev> scanner is tabulated, and the percentage of variation is shown to be within the tolerance limits at all kVp at 5 mAs for <abbrev xlink:title="Computed tomography" id="ABBRID0E4XAE">CT</abbrev>, as illustrated in <bold>Table <xref ref-type="table" rid="T2">2</xref></bold>. The measured operating potential for 80, 110, and 130 kVp was tabulated, and the variation was within the tolerance limit provided.</p>
      </sec>
      <sec sec-type="High contrast resolution" id="SECID0EHYAE">
        <title>High contrast resolution</title>
        <p>Two major categories of measuring techniques exist: the examination of the point spread function, typically through the determination of the modulation transfer function (<abbrev xlink:title="modulation transfer function" id="ABBRID0ENYAE">MTF</abbrev>). The resolution is expressed as the spatial frequency (in line pairs per centimeter) at which the modulation falls to 50%, 10%, or 2% <abbrev xlink:title="modulation transfer function" id="ABBRID0ERYAE">MTF</abbrev>, depending on whether it is determined through objective analysis or visual evaluation of images of a resolution bar phantom. The use of a body and head phantom facilitates the confirmation of slice thickness.<sup>[<xref ref-type="bibr" rid="B14">14</xref>]</sup> The stationary mode option in the service mode is employed to fix the exposure factors at a zero-degree angle. The phantom is positioned centrally on the table and exposed using a low kVp of 80 mAs and 2 mAs at intervals of 10 cm.</p>
      </sec>
      <sec sec-type="kVp accuracy" id="SECID0E3YAE">
        <title>kVp accuracy</title>
        <p>kVp accuracy is tested across the typical range (80, 110, and 130 kVp) at 100, 200, and 300 mA stations in localization scan mode with the tube above the table. Scans start 10 cm from one side of the detector and end 10 cm from the gantry, ensuring simultaneous irradiation. The kVp tolerance is ±2 kVp. Measured operating potentials for 80, 110, and 130 kVp are tabulated, with variations within the tolerance limit <bold>(Table <xref ref-type="table" rid="T2">2</xref>)</bold>. The percentage deviation across four <abbrev xlink:title="Computed tomography" id="ABBRID0EJZAE">CT</abbrev> scanners is 8%, 7.3%, 7.8%, and 5%, respectively, within the ±10% tolerance.</p>
      </sec>
      <sec sec-type="The timer accuracy" id="SECID0ENZAE">
        <title>The timer accuracy</title>
        <p>The loading time must not deviate from the selected value by more than 10% +1 msec. Timer accuracy for exposure times of 1.0, 1.5, and 2.0 seconds across four <abbrev xlink:title="Computed tomography" id="ABBRID0ETZAE">CT</abbrev> scanners showed errors of 0.8%, 0.4%, 0.6%, and 0.75%, respectively, within acceptable limits.<sup>[<xref ref-type="bibr" rid="B15">15</xref>]</sup></p>
      </sec>
      <sec sec-type="mR/mAs linearity" id="SECID0E4ZAE">
        <title>mR/mAs linearity</title>
        <p>The mR/mAs linearity is verified at 110 kVp for a 1.0-second exposure at 100, 150, and 200 mA stations using the Piranha. The coefficient of linearity (<abbrev xlink:title="coefficient of linearity" id="ABBRID0ED1AE">COL</abbrev>) is calculated as:</p>
        <p>
          <mml:math id="M2">
            <mml:mtext>Coefficient of linearity</mml:mtext>
            <mml:mo>(</mml:mo>
            <mml:mi>COL</mml:mi>
            <mml:mo>)</mml:mo>
            <mml:mo>=</mml:mo>
            <mml:mfrac>
              <mml:mrow>
                <mml:msub>
                  <mml:mi>X</mml:mi>
                  <mml:mo>max</mml:mo>
                </mml:msub>
                <mml:mo>−</mml:mo>
                <mml:msub>
                  <mml:mi>X</mml:mi>
                  <mml:mo>min</mml:mo>
                </mml:msub>
              </mml:mrow>
              <mml:mrow>
                <mml:msub>
                  <mml:mi>X</mml:mi>
                  <mml:mo>max</mml:mo>
                </mml:msub>
                <mml:mo>+</mml:mo>
                <mml:msub>
                  <mml:mi>X</mml:mi>
                  <mml:mo>min</mml:mo>
                </mml:msub>
              </mml:mrow>
            </mml:mfrac>
          </mml:math>
        </p>
        <p>The X-ray tube current must not deviate by more than 10% +0.2 mAs.</p>
      </sec>
      <sec sec-type="Output consistency" id="SECID0ED2AE">
        <title>Output consistency</title>
        <p>Output consistency is verified at 80, 110, and 130 kVp with 100 mAs and a 1-second exposure using the Piranha. The coefficient of variation (<abbrev xlink:title="coefficient of variation" id="ABBRID0EJ2AE">COV</abbrev>) is calculated as:</p>
        <p>COV=X-1 (å[Xi -X]2/n-1)½</p>
        <p>*Tolerance in COV: ≤0.05</p>
        <p>All COV values are below the tolerance of ≤0.05 for all CTs.</p>
      </sec>
      <sec sec-type="Total filtration" id="SECID0EQ2AE">
        <title>Total filtration</title>
        <p>Total filtration is measured using the Piranha 657 at 110 kVp, 50 mA, and a 1-second exposure. For machine A, total filtration was 7.2 mm Al at 130 kVp, 100 mA, with a half-value layer (<abbrev xlink:title="half-value layer" id="ABBRID0EW2AE">HVL</abbrev>) of 6.96 mm Al. For machine B, it was 5.78 mm Al at 110 kVp, 100 mAs, with an <abbrev xlink:title="half-value layer" id="ABBRID0E12AE">HVL</abbrev> of 4.86 mm Al. Tolerance for total filtration was 2 mm Al at 100 kV and 2.5 mm Al for machines A, B, C, and D, all within limits. The percentage variations between two CTs were −3.4%, −18.9%, −7.4%, and 8.4%, respectively.<sup>[<xref ref-type="bibr" rid="B16">16</xref>]</sup></p>
      </sec>
      <sec sec-type="CT number accuracy" id="SECID0EE3AE">
        <title><abbrev xlink:title="Computed tomography" id="ABBRID0EJ3AE">CT</abbrev> number accuracy</title>
        <p>The reconstruction algorithm is responsible for generating the <abbrev xlink:title="Computed tomography" id="ABBRID0EP3AE">CT</abbrev> number, which is associated with the attenuation coefficient. The <abbrev xlink:title="Computed tomography" id="ABBRID0ET3AE">CT</abbrev> number range is from −1000 for air to +1000 for bone, with the value for water set at 0. Numbers that have been normalized in this manner are referred to as Hounsfield Units (HU).<sup>[<xref ref-type="bibr" rid="B17">17</xref>]</sup></p>
        <p>
          <mml:math id="M3">
            <mml:mi>HU</mml:mi>
            <mml:mo>=</mml:mo>
            <mml:mfrac>
              <mml:mrow>
                <mml:msub>
                  <mml:mi>μ</mml:mi>
                  <mml:mi>m</mml:mi>
                </mml:msub>
                <mml:mo>−</mml:mo>
                <mml:msub>
                  <mml:mi>μ</mml:mi>
                  <mml:mi>w</mml:mi>
                </mml:msub>
              </mml:mrow>
              <mml:msub>
                <mml:mi>μ</mml:mi>
                <mml:mi>w</mml:mi>
              </mml:msub>
            </mml:mfrac>
            <mml:mo>×</mml:mo>
            <mml:mn>1000</mml:mn>
          </mml:math>
        </p>
        <p>where µ<sub>m</sub> is the material’s attenuation coefficient and µ<sub>w</sub> is the attenuation coefficient of water.</p>
        <p>The <abbrev xlink:title="Computed tomography" id="ABBRID0E14AE">CT</abbrev> number accuracy for water is verified using a water phantom filled with distilled water.<sup>[<xref ref-type="bibr" rid="B18">18</xref>]</sup> The water phantom is scanned at 130 kVp, 5 mm slice thickness, 50 mA, and a scan time of 68.29 seconds. The solid water phantom (30×30×30 cm<sup>3</sup>) is positioned on the couch, and readings are recorded at various positions.<sup>[<xref ref-type="bibr" rid="B19">19</xref>]</sup></p>
        <p>The <abbrev xlink:title="Computed tomography" id="ABBRID0EP5AE">CT</abbrev> number accuracy test for water was carried out on four machines, and the results were tabulated. The tolerance level was ±5 HU, with a maximum of +6 HU and a minimum of −2 HU, all values being within the tolerance level.</p>
      </sec>
      <sec sec-type="CTDI measurement" id="SECID0ET5AE">
        <title><abbrev xlink:title="computed tomography dose index" id="ABBRID0EY5AE">CTDI</abbrev> measurement</title>
        <p>The measurements of central and peripheral <abbrev xlink:title="computed tomography dose index" id="ABBRID0E55AE">CTDI</abbrev> using a pencil beam ion chamber in head and body phantoms are shown in <bold>Table <xref ref-type="table" rid="T2">2</xref></bold>. The percentage of variation was within the tolerance limit for all CTs. Weighted <abbrev xlink:title="computed tomography dose index" id="ABBRID0EI6AE">CTDI</abbrev> (<abbrev xlink:title="Weighted CTDI" id="ABBRID0EM6AE">CTDIw</abbrev>) is equal to 1/3 <abbrev xlink:title="computed tomography dose index" id="ABBRID0EQ6AE">CTDI</abbrev><sub>central</sub> + 2/3 <abbrev xlink:title="computed tomography dose index" id="ABBRID0EV6AE">CTDI</abbrev><sub>peripheral</sub>. Stated CTDIW (head phantom): 6.55 mGy at 100 mAs, CTDIW (head phantom): 2.66 mGy at 100 mAs, measured CTDIW (head phantom): 8.661 mGy at 100 mAs, CTDIW (head phantom): 3.635 mGy at 100 mAs.</p>
        <table-wrap id="T3" position="float" orientation="portrait">
          <label>Table 3.</label>
          <caption>
            <p>The radiation protection of computed tomography installation (Simulation <abbrev xlink:title="Computed tomography" id="ABBRID0EEAAG">CT</abbrev>)</p>
          </caption>
          <table id="TID0EJGAG" rules="all">
            <tbody>
              <tr>
                <td rowspan="1" colspan="1">
                  <bold>S. No</bold>
                </td>
                <td rowspan="1" colspan="1">
                  <bold>Location</bold>
                </td>
                <td rowspan="1" colspan="1"><bold>Survey meter reading </bold> µsv/hr</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">1</td>
                <td rowspan="1" colspan="1">Lead lined door</td>
                <td rowspan="1" colspan="1">1.07</td>
              </tr>
              <tr>
                <td rowspan="3" colspan="1">2</td>
                <td rowspan="1" colspan="1">Lead glass viewing window A</td>
                <td rowspan="1" colspan="1">2.96</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">B</td>
                <td rowspan="1" colspan="1">2.80</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">C</td>
                <td rowspan="1" colspan="1">3.10</td>
              </tr>
              <tr>
                <td rowspan="3" colspan="1">3</td>
                <td rowspan="1" colspan="1">Control console A</td>
                <td rowspan="1" colspan="1">0.08</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">B</td>
                <td rowspan="1" colspan="1">0.08</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">C</td>
                <td rowspan="1" colspan="1">0.17</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">4</td>
                <td rowspan="1" colspan="1">Main entrance door</td>
                <td rowspan="1" colspan="1">0.37</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">5</td>
                <td rowspan="1" colspan="1">Linac server room</td>
                <td rowspan="1" colspan="1">0.10</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">6</td>
                <td rowspan="1" colspan="1">TPS room</td>
                <td rowspan="1" colspan="1">0.10</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">7</td>
                <td rowspan="1" colspan="1">Patient waiting area</td>
                <td rowspan="1" colspan="1">0.05</td>
              </tr>
              <tr>
                <td rowspan="1" colspan="1">8</td>
                <td rowspan="1" colspan="1">Open space and road</td>
                <td rowspan="1" colspan="1">0.08</td>
              </tr>
            </tbody>
          </table>
        </table-wrap>
      </sec>
      <sec sec-type="Radiation protection survey of CT" id="SECID0ETEAG">
        <title>Radiation protection survey of CT</title>
        <p>Radiation leakage from the X-ray tube casing at 1 meter (simulation <abbrev xlink:title="Computed tomography" id="ABBRID0E4EAG">CT</abbrev> and three diagnostic CTs) was measured at 130 kVp, 1 mAs, and 15 seconds. Leakage levels were 54 mR/hr (front cathode), 23.1 mR/hr (back anode), 1.14 mR/hr (left), and 1.94 mR/hr (right). Maximum leakage for a 500 mA/min workload was 4.5 mR/hr, below the recommended limit of ≤115 mR/hr. For a 180 mA/min workload at 110 kVp, 100 mA, and 1 second, leakage levels were 9.70 mR/hr (front cathode), 6.24 mR/hr (back anode), 2.31 mR/hr (left), and 6.12 mR/hr (right), with a maximum of 0.808 mR/hr, within limits.</p>
      </sec>
      <sec sec-type="Operating potential" id="SECID0EBFAG">
        <title>Operating potential</title>
        <p>kVp 110 kV, 100 mA, exposure time 1 second, workload 180 mA min/hr. At a distance of one meter from the machine, the measured radiation leakage was 9.70 mR/hr for the cathode, 6.24 mR/hr for the anode, 2.31 mR/hr for the left side, and 6.12 mR/hr for the right side.</p>
        <p>kVp 130 kV, 500 mA, slice thickness 5 mm, exposure time 1 second. High leakage = 500 mA min in one hour × 9.70 mR/hr. The high leakage 60×100 mA was used for measurement.</p>
        <p>As a result, we found that the highest radiation leakage at 1 meter from the casing for a workload of 180 mA min in an hour was 0.808 mR. The recommended upper limit leakage radiation level at 1 meter from the focus should be ≤115 mR in one hour.</p>
      </sec>
      <sec sec-type="Radiation protection of computed tomography installation (B, C, D)" id="SECID0EIFAG">
        <title>Radiation protection of computed tomography installation (B, C, D)</title>
        <p>Detail of radiation protection survey with mA: 100 kV: 110 and exposure time: 1.0. The maximum radiation level/week (mR/wk) = 150 mA min/week was 0.8, the high radiation level/week (mR/wk) was 0.02, and the permissible limit of radiation for a radiation worker was 20 mSv/year (40 mR/week) and for a member of the public was 1 mSv/year (2 mR/week). The radiation protection survey reading of the computed tomography installation was analyzed, and the percentage of variations was within the tolerance limits mentioned in <bold>Table <xref ref-type="table" rid="T2">2</xref></bold>.</p>
      </sec>
      <sec sec-type="Diagnostic practical cases and real patient and dynamic investigations" id="SECID0EUFAG">
        <title>Diagnostic practical cases and real patient and dynamic investigations.</title>
        <p>A considerable number of artifacts from the early days of <abbrev xlink:title="Computed tomography" id="ABBRID0E1FAG">CT</abbrev> have been significantly reduced, yet some have persisted. The advent of new technologies has led to the emergence of more fully characterized artifacts, as Quality Assurance (<abbrev xlink:title="quality assurance" id="ABBRID0E5FAG">QA</abbrev>) procedures are not conducted on an annual basis. <bold>Figs <xref ref-type="fig" rid="F4">4</xref>, <xref ref-type="fig" rid="F5">5</xref></bold>, and <bold>6</bold> show some of the examples where artifacts were found in our department prior to <abbrev xlink:title="quality assurance" id="ABBRID0EOGAG">QA</abbrev>.</p>
        <fig id="F4" position="float" orientation="portrait">
          <object-id content-type="arpha">4D1A67CB-7FD7-50A2-8E9F-57BF591DCFC7</object-id>
          <label>Figure 4.</label>
          <caption>
            <p>Artifact detected thorax images due to circular high-density areas in both thighs indicate metallic implants, likely orthopedic hardware. Surrounding them are dark streaks, typical of beam hardening and metal-induced streak artifacts surround them.</p>
          </caption>
          <graphic xlink:href="foliamedica-67-5-e153885-g004.jpg" position="float" orientation="portrait" xlink:type="simple" id="oo_1446613.jpg">
            <uri content-type="original_file">https://binary.pensoft.net/fig/1446613</uri>
          </graphic>
        </fig>
        <fig id="F5" position="float" orientation="portrait">
          <object-id content-type="arpha">9D7F832F-2D60-57D1-A8E9-C96C8351CF4F</object-id>
          <label>Figure 5.</label>
          <caption>
            <p>Streak artifact detected abdomen images due to <abbrev xlink:title="Computed tomography" id="ABBRID0EGHAG">CT</abbrev> contrast artifacts are image distortions caused by contrast media, appearing as streaks, blurring, or false enhancement. They can affect diagnosis but are reduced with proper timing, technique, and corrections.</p>
          </caption>
          <graphic xlink:href="foliamedica-67-5-e153885-g005.jpg" position="float" orientation="portrait" xlink:type="simple" id="oo_1446614.jpg">
            <uri content-type="original_file">https://binary.pensoft.net/fig/1446614</uri>
          </graphic>
        </fig>
        <fig id="F6" position="float" orientation="portrait">
          <object-id content-type="arpha">097C3545-B854-561E-A531-3A9956FB7905</object-id>
          <label>Figure 6.</label>
          <caption>
            <p>Streak artifact detected in brain images due to radiation dosage factor (also known as beam hardening artifact), most notably in the posterior fossa and periventricular area in some coronal and sagittal views. Electronic noise detected on impact with image noise.</p>
          </caption>
          <graphic xlink:href="foliamedica-67-5-e153885-g006.jpg" position="float" orientation="portrait" xlink:type="simple" id="oo_1446615.jpg">
            <uri content-type="original_file">https://binary.pensoft.net/fig/1446615</uri>
          </graphic>
        </fig>
      </sec>
    </sec>
    <sec sec-type="Discussion" id="SECID0E2HAG">
      <title>Discussion</title>
      <p>A series of mechanical and radiation tests for <abbrev xlink:title="Computed tomography" id="ABBRID0EBIAG">CT</abbrev> simulators and diagnostic <abbrev xlink:title="Computed tomography" id="ABBRID0EFIAG">CT</abbrev> have been conducted in accordance with the <abbrev xlink:title="Atomic Energy Regulatory Board" id="ABBRID0EJIAG">AERB</abbrev> protocol, and the results were found to be within the acceptable limits. The maximum deviation between the set slice thickness and measured slice thickness for different slice thicknesses of 1.4 mm and 5 mm was −0.2 mm. The maximum deviation between the set and measured kVp for 80, 110, and 130 kVp for different mA stations of 100, 150, and 200 mA was −17 kVp. The maximum percentage of timer error was 0.8% for 1.0 second. The same value has been repeated in the diagnostic <abbrev xlink:title="Computed tomography" id="ABBRID0ENIAG">CT</abbrev>.</p>
      <p>The three parts address patient exposure: radiographic positioning, loading parameters, and entrance-skin exposure. The guidelines in each section are intended to improve the diagnostic quality of radiographs while reducing patient exposure. To minimize patient exposure, the set time and measurement time should be the same. Exposure is critical for timer accuracy tests since it can prevent overexposure to the subject.</p>
      <p>The coefficient of linearity for 100, 150, and 200 mA stations was 0.0014, which is much less than the tolerance value of 0.1. The coefficient of variation of X-ray output consistency for 80, 110, and 130 kVp was 0.0013, 0.0005, and 0.0011, respectively. Values were much less than the tolerance value of 0.05. The total filtration for 110 kVp was well within the tolerance limit of 7.2 mm of Al equivalent.</p>
      <p>The <abbrev xlink:title="computed tomography dose index" id="ABBRID0EVIAG">CTDI</abbrev> values displayed in the console for the head and body phantom match well with the measured <abbrev xlink:title="computed tomography dose index" id="ABBRID0EZIAG">CTDI</abbrev>, and they are well within the acceptable maximum tolerance limit of ±40% of the stated values, and the same result has been repeated for diagnostic <abbrev xlink:title="Computed tomography" id="ABBRID0E4IAG">CT</abbrev>.<sup>[<xref ref-type="bibr" rid="B20">20</xref>]</sup></p>
      <p>The resolution for high-contrast resolution images is 0.5 lp/mm, but the low-contrast resolution image resolved a 3 mm line pair at 1% contrast difference. The maximum leakage level of X-ray tube housing at 1 m distance is 4.5 mR/hr. The radiation protection survey measurements confirm the integrity of radiation safety of the installation, as all the values are within the acceptable limits of 10 µSv hours for the area occupied by the radiation worker and for the general public, respectively.</p>
      <p>All examination procedures must be optimized for both image quality and radiation dose, and both must undergo routine quality assurance assessment.<sup>[<xref ref-type="bibr" rid="B21">21</xref>]</sup> The goal of this research was to assess several <abbrev xlink:title="Computed tomography" id="ABBRID0ERJAG">CT</abbrev> machines. Our findings show that Piranha dosimetry is consistent over time and that there is only minimal acceptable variation in the results of four <abbrev xlink:title="Computed tomography" id="ABBRID0EVJAG">CT</abbrev> scans, making Piranha a useful tool for quality assurance.</p>
      <p>The use of HU measurements for diagnostic purposes is discussed in a number of studies, along with recommendations for tissue features, tissue differentiation, and threshold values for malignant and benign masses, such as in the adrenal glands. The measurements of HU for various item densities between various <abbrev xlink:title="Computed tomography" id="ABBRID0E2JAG">CT</abbrev> scanners, as well as over time for the same scanner, demonstrate a significant difference in our results.</p>
      <p>Radiologists should be aware of this as it pertains to the use of HU for tissue. We suggest that HU may be used cautiously for diagnostic reasons based on the findings of our investigation. The HU is crucial for simulation <abbrev xlink:title="Computed tomography" id="ABBRID0EBKAG">CT</abbrev>, and the radiation oncologist should beware of lead marker or attenuation factor-induced bone marker artifact.<sup>[<xref ref-type="bibr" rid="B22">22</xref>]</sup> As a result, the artifact will have an impact on treatment planning. To solve this issue, the lead marker should be kept close to the patient, and the marker size should be minimal.</p>
      <p>Diagnostic <abbrev xlink:title="Computed tomography" id="ABBRID0EOKAG">CT</abbrev> scanners typically only have a cradle-shaped couch top; the tabletop is cup-shaped to fit the <abbrev xlink:title="Computed tomography" id="ABBRID0ESKAG">CT</abbrev> scanner gantry’s circular hole. A flat tabletop that closely resembles the tabletop geometry of the treatment machine is necessary for scanners used for <abbrev xlink:title="Computed tomography" id="ABBRID0EWKAG">CT</abbrev> simulation. The flat tabletop can be either an overlay installed on top of the existing table’s cradle or an insert that slips into the cradle of the current table. In terms of treatment setup precision, the <abbrev xlink:title="Computed tomography" id="ABBRID0E1KAG">CT</abbrev> scanner and the therapy device are directly connected by the tabletop. Poor patient position reproducibility on the treatment equipment results from errors in the scanner tabletop geometry. Vertical and longitudinal movement faults can lead to improper marking of the patient’s skin in relation to the calculated treatment isocenter. Improper table indexing can also lead to image spatial distortions. Our investigation found that the patient dose and image quality were unaffected by the <abbrev xlink:title="Computed tomography" id="ABBRID0E5KAG">CT</abbrev> table’s flat or curved surfaces. Scanning techniques were as similar as possible for all scanners. The same Piranha device was, as much as possible, used for measurements on the same <abbrev xlink:title="Computed tomography" id="ABBRID0ECLAG">CT</abbrev> scanners over time, although different dosimetry was applied for measurements on other <abbrev xlink:title="Computed tomography" id="ABBRID0EGLAG">CT</abbrev> scanners. However, according to the worldwide organization for radiation safety, there are no discrepancies between four centers using the same dosimetry.</p>
      <p>The artifact absorbed in our <abbrev xlink:title="Computed tomography" id="ABBRID0EMLAG">CT</abbrev> equipment is technically advanced and has to be further improved. To avoid image artifacts, <abbrev xlink:title="quality assurance" id="ABBRID0EQLAG">QA</abbrev> adherence has to be followed.<sup>[<xref ref-type="bibr" rid="B23">23</xref>]</sup> Following the occurrence of the streak artifact, beam hardening artifact, ring artifact, and metal artifact due to parameter and technical issues, we conducted a quality control inspection on <abbrev xlink:title="Computed tomography" id="ABBRID0E2LAG">CT</abbrev> equipment.<sup>[<xref ref-type="bibr" rid="B24">24</xref>]</sup> The result was deemed to be an acceptable limit according to the <abbrev xlink:title="Atomic Energy Regulatory Board" id="ABBRID0EGMAG">AERB</abbrev> protocol values.<sup>[<xref ref-type="bibr" rid="B25">25</xref>]</sup></p>
      <p>According to <abbrev xlink:title="Atomic Energy Regulatory Board" id="ABBRID0ESMAG">AERB</abbrev> and NABH Quality Assurance for Diagnostic Radiology Machines, two years is sufficient. However, daily <abbrev xlink:title="quality assurance" id="ABBRID0EWMAG">QA</abbrev> checks are required for radiotherapy <abbrev xlink:title="Computed tomography" id="ABBRID0E1MAG">CT</abbrev> machines since <abbrev xlink:title="Computed tomography" id="ABBRID0E5MAG">CT</abbrev> positioning accuracy and HU units are essential for treatment accuracy.<sup>[<xref ref-type="bibr" rid="B26">26</xref>]</sup></p>
    </sec>
    <sec sec-type="Conclusion" id="SECID0EINAG">
      <title>Conclusion</title>
      <p>The result of the mechanical and radiation tests of one <abbrev xlink:title="Computed tomography" id="ABBRID0EONAG">CT</abbrev> simulator and three diagnostic CTs shows that Siemens Emotion Duo and GE can be used clinically to obtain optimized images with minimum radiation dose to the patient. It is mandatory to carry out periodic <abbrev xlink:title="Computed tomography" id="ABBRID0ESNAG">CT</abbrev><abbrev xlink:title="quality assurance" id="ABBRID0EWNAG">QA</abbrev> as per the national and international <abbrev xlink:title="Computed tomography" id="ABBRID0E1NAG">CT</abbrev><abbrev xlink:title="quality assurance" id="ABBRID0E5NAG">QA</abbrev> protocols to ensure good quality of image.</p>
    </sec>
    <sec sec-type="Funding" id="SECID0ECOAG">
      <title>Funding</title>
      <p>Authors have no funding to report.</p>
    </sec>
    <sec sec-type="Competing interests" id="SECID0EHOAG">
      <title>Competing interests</title>
      <p>Тhe authors have declared that no competing interests exist.</p>
    </sec>
  </body>
  <back>
    <ack>
      <title>Acknowledgments</title>
      <p>The authors have no support to report.</p>
    </ack>
    <ref-list>
      <title>References</title>
      <ref id="B1">
        <mixed-citation xlink:type="simple">1. Nocum DJ, Robinson J, Reed W. The role of quality improvement in radiography. J Med Radiat Sci 2021; 68(3):214–6.</mixed-citation>
      </ref>
      <ref id="B2">
        <mixed-citation xlink:type="simple">2. Kalra MK, Maher MM, Toth TL, et al. Strategies for CT radiation dose optimization. Radiology 2004; 230(3):619–28. doi: <ext-link xlink:type="simple" ext-link-type="doi" xlink:href="10.1148/radiol.2303021966">10.1148/radiol.2303021966</ext-link></mixed-citation>
      </ref>
      <ref id="B3">
        <mixed-citation xlink:type="simple">3. Matsubara K, Ichikawa K, Murasaki Y, et al. Accuracy of measuring half‐and quarter‐value layers and appropriate aperture width of a convenient method using a lead‐covered case in X‐ray computed tomography. J Appl Clin Med Phys 2014; 15(1):309–16.</mixed-citation>
      </ref>
      <ref id="B4">
        <mixed-citation xlink:type="simple">4. Kang EJ. Clinical applications of wide-detector CT scanners for cardiothoracic imaging: an update. Korean J Radiol 2019; 20(12):1583.</mixed-citation>
      </ref>
      <ref id="B5">
        <mixed-citation xlink:type="simple">5. Schmidgunst C, Ritter D, Lang E. Calibration model of a dual gain flat panel detector for 2D and 3D x‐ray imaging. Med Phys 2007; 34(9):3649–64.</mixed-citation>
      </ref>
      <ref id="B6">
        <mixed-citation xlink:type="simple">6. McCollough CH, Leng S, Yu L, et al. CT dose index and patient dose: they are not the same thing. Radiology 2011; 259(2):311–6.</mixed-citation>
      </ref>
      <ref id="B7">
        <mixed-citation xlink:type="simple">7. Bishnu M, Vijiyan V. In-air calibration of an HDR <sup>192</sup>Ir brachytherapy source using therapy ion chambers. J Cancer Res Ther 2005; 1(4):213–20.</mixed-citation>
      </ref>
      <ref id="B8">
        <mixed-citation xlink:type="simple">8. Descamps C, Gonzalez M, Garrigo E, et al. Measurements of the dose delivered during CT exams using AAPM Task Group Report No. 111. J Appl Clin Med Phys 2012; 13(6):293–302.</mixed-citation>
      </ref>
      <ref id="B9">
        <mixed-citation xlink:type="simple">9. Rowshanfarzad P, Riis HL, Zimmermann SJ, et al. A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation. British J Radiol 2015; 88(1051):20140581.</mixed-citation>
      </ref>
      <ref id="B10">
        <mixed-citation xlink:type="simple">10. DeWerd LA, Micka JA, Laird RW, et al. The effect of spectra on calibration and measurement with mammographic ionization chambers. Med Phys 2002; 29(11):2649–54.</mixed-citation>
      </ref>
      <ref id="B11">
        <mixed-citation xlink:type="simple">11. Racine D, Viry A, Becce F, et al. Objective comparison of high‐contrast spatial resolution and low‐contrast detectability for various clinical protocols on multiple CT scanners. Med Phys 2017; 44(9):e153–63.</mixed-citation>
      </ref>
      <ref id="B12">
        <mixed-citation xlink:type="simple">12. Lin E, Alessio A. What are the basic concepts of temporal, contrast, and spatial resolution in cardiac CT? J Cardiovasc Comput Tomogr 2009; 3(6):403–8.</mixed-citation>
      </ref>
      <ref id="B13">
        <mixed-citation xlink:type="simple">13. Fitzgerald PF, Colborn RE, Edic PM, et al. CT image contrast of high-Z elements: phantom imaging studies and clinical implications. Radiology 2016; 278(3):723–33.</mixed-citation>
      </ref>
      <ref id="B14">
        <mixed-citation xlink:type="simple">14. Mkimel M, Mesradi MR, El Baydaoui R, et al. Assessment of computed tomography dose index (CTDI) using the platform GEANT4/GATE. Perspect Sci 2019; 12:100405.</mixed-citation>
      </ref>
      <ref id="B15">
        <mixed-citation xlink:type="simple">15. Szarmach A, Piskunowicz M, Świętoń D, et al. Radiation safety awareness among medical staff. Polish J Radiol 2015; 80:57.</mixed-citation>
      </ref>
      <ref id="B16">
        <mixed-citation xlink:type="simple">16. Greffier J, Pereira F, Hamard A, et al. Effect of tin filter-based spectral shaping CT on image quality and radiation dose for routine use on ultralow-dose CT protocols: a phantom study. Diagn Interv Imaging 2020; 101(6):373–81.</mixed-citation>
      </ref>
      <ref id="B17">
        <mixed-citation xlink:type="simple">17. Hess J, Bender ML, Schilling JG. Evolution of the ratio of strontium-87 to strontium-86 in seawater from Cretaceous to present. Science 1986; 231(4741):979–84.</mixed-citation>
      </ref>
      <ref id="B18">
        <mixed-citation xlink:type="simple">18. Pang T, Yang B, Liu X, et al. Investigation of absolute dose calibration accuracy for TomoTherapy using real water. J Appl Clin Med Phys 2021; 22(6):139–45.</mixed-citation>
      </ref>
      <ref id="B19">
        <mixed-citation xlink:type="simple">19. Clausen M, Khachonkham S, Gruber S, et al. Phantom design and dosimetric characterization for multiple simultaneous cell irradiations with active pencil beam scanning. Radiat Environ Biophys 2019; 58:563–73.</mixed-citation>
      </ref>
      <ref id="B20">
        <mixed-citation xlink:type="simple">20. Saravanakumar A, Vaideki K, Govindarajan KN, et al. Establishment of diagnostic reference levels in computed tomography for select procedures in Pudhuchery, India. J Med Phys 2014; 39(1):50–5.</mixed-citation>
      </ref>
      <ref id="B21">
        <mixed-citation xlink:type="simple">21. Schumacher A, Sihn W. Development of a monitoring system for implementation of industrial digitalization and automation using 143 key performance indicators. Procedia CIRP 2020; 93:1310–5.</mixed-citation>
      </ref>
      <ref id="B22">
        <mixed-citation xlink:type="simple">22. Liu Y, Lei Y, Wang Y, et al. Evaluation of a deep learning-based pelvic synthetic CT generation technique for MRI-based prostate proton treatment planning. Phys Med Biol 2019; 64(20):205022.</mixed-citation>
      </ref>
      <ref id="B23">
        <mixed-citation xlink:type="simple">23. Selles M, Van Osch JA, Maas M, et al. Advances in metal artifact reduction in CT images: A review of traditional and novel metal artifact reduction techniques. Eur J Radiol 2024; 170:111276.</mixed-citation>
      </ref>
      <ref id="B24">
        <mixed-citation xlink:type="simple">24. Boas FE, Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med 2012; 4(2):229–40.</mixed-citation>
      </ref>
      <ref id="B25">
        <mixed-citation xlink:type="simple">25. Tschauner S, Marterer R, Nagy E, et al. Experiences with image quality and radiation dose of cone beam computed tomography (CBCT) and multidetector computed tomography (MDCT) in pediatric extremity trauma. Skeletal Radiology 2020; 49:1939–49.</mixed-citation>
      </ref>
      <ref id="B26">
        <mixed-citation xlink:type="simple">26. Bannas P, Habermann CR, Jung C, et al. Diagnostic accuracy of state-of-the-art MDCT scanners without gantry tilt in patients with oral and oropharyngeal cancer. Eur J Radiol 2012; 81(12):3947–52.</mixed-citation>
      </ref>
    </ref-list>
  </back>
</article>