Simeonov N, Stefanova V (2023) Micromorphology and analysis of dentin surfaces after preparation with Er:YAG laser and application of self-etch adhesive system with 10-MDP. A confocal laser scanning microscope study. Folia Medica 65(3): 453-459. https://doi.org/10.3897/folmed.65.e76606
|
Aim: The purpose of this study was to analyze and compare the micromorphology of the hybrid layer and the dentinal surface when the dentin was prepared with an Er:YAG laser or burs, and to find if there was any difference if the time of application of the self-etching primer on the dentin prepared with an Er:YAG laser was doubled.
Materials and methods: Fifteen freshly extracted human teeth with preserved crowns were selected. Three cavities were prepared on each extracted tooth. Two of the cavities were prepared with an Er:YAG dental laser, and the third with burs. A self-etching adhesive system was applied. The primer of the adhesive was labeled with Rhodamine B, and the bond was labeled with fluorescein prior to application. The teeth were sectioned mesiodistally through the cavities and the cavity of each sample was examined using a confocal laser scanning microscope. The infiltration and micromorphology were determined qualitatively on ninety samples.
Results: The dentinal surface in the samples prepared with burs was smoother and with shorter resin tags than the surface prepared with an Er:YAG laser. When using an Er:YAG laser for preparation, crater-like irregularities of the surface were observed. In five of the samples prepared with burs, no resin tags were found. We could not detect any difference in the hybrid layer when the time of application of the adhesive system was increased.
Conclusions: The clinical significance of the tag length and quality, as well as the infiltration ratio, needs to be further studied.
confocal laser scanning microscope, Er:YAG, hybrid layer, MDP, resin tags
Composite materials in restorative dentistry require the use of adhesive systems. These systems create a micromechanical bond between dentin/enamel and the composite material.[
Lasers are high-tech tools that have been successfully used in dentistry for numerous manipulations.[
Confocal laser scanning microscope (CLSM) allows in-depth observation of the surface, which makes it a good choice when studying the hybrid layer and its penetration.[
In the present study, we aimed to analyze and compare the micromorphological structure of the hybrid layer and the dentinal surface when the dentin was prepared with an Er:YAG laser or with burs, and to find if there would be any difference if the application time of the self-etching primer on the dentin prepared with an Er:YAG laser increased two times.
Using visual inspection, we selected 15 freshly extracted human teeth with preserved crowns and no carious and/or non-carious lesions and fillings; these teeth were stored in a 0.2% aqueous-alcoholic solution of thymol at a temperature of 2–6°C. In this study, we used the Clearfil SE Bond 2 adhesive system (Kuraray, Japan) which contained 10-MDP. The primer was mixed with Rhodamine B and the bond – with fluorescein, with both dyes used at a concentration of 0.1%. Their emission peaks are different, which makes it possible to examine them simultaneously with CLSM.
Three cavities were prepared in each extracted tooth until the middle zone of the dentin was reached – one on the occlusal surface, one on the mesial surface, and one on the distal surface. All cavities were box-shaped having five walls and one open surface. Random two cavities were prepared with an Er:YAG dental laser (LiteTouch, Syneron, Israel) with the manufacturer’s recommended sapphire tips and settings. We used burs in the preparation of the third cavity.
The prepared surfaces were wiped thoroughly with a cotton pellet soaked in 3% hydrogen peroxide, then with a cotton pellet soaked in 70% alcohol, and finally dried gently with air from the dental unit. The cavities prepared with the Er:YAG dental laser were randomly divided into two groups. The adhesive system was applied to the cavities of group 1 in strict accordance with the manufacturer’s instructions (self-etching primer for 20 s, air-drying, bonding, and light-curing); the cavities of group 2 were prepared by applying the adhesive system twice as long as specified by the manufacturer (self-etching primer for 40 s, air-drying, bonding, and light-curing). The same adhesive system was applied to the cavities prepared with burs according to the manufacturer’s instructions (group 3 – controls). Using a diamond saw, all teeth were split along the longitudinal axis in the mesiodistal direction simultaneously through the three cavities.
The adhesive system and the prepared dentin in the area of shear were examined with a confocal laser-scanning microscope (Leica TCS SPE DM2500) with optical magnification 10× and the possibility of additional digital magnification using software Leica Microsystems LAS AF – TCS SPE. The size of the observed field at 10× magnification was 1100×1100 μm with the possibility of changing the resolution. The areas of interest were scanned and saved as digital images.
The infiltration and the micromorphology were determined qualitatively on ninety samples. Only the fluorescent dyes added to the adhesive system were visualized on the images obtained by CLSM. The primer was displayed in red and the bond – in green. When a superimposed image of the two dyes was created, the areas where they coincided were colored in yellow.
What are visualized are the primer and the bond. An image of the relevant area is seen as a strip that depicts the adhesive system on the dentinal surface and resin-tags, which are in the form of cones entering the dentinal tubules. In five of the samples prepared with burs, no resin-tags were found.
Analysis of the images in group 1 (Fig.
The dentinal surface prepared with burs appears to be smoother than the dentinal surface prepared with the Er:YAG laser. It is noteworthy that the images of the fluorescent primer do not completely match the areas of the fluorescent adhesive. This is very clearly seen in the superimposed image (Figs
We could not find a significant difference in the hybrid layer when the time of application of the adhesive system was increased (Figs
In the control group, both the primer and the bond created a dense, uniform layer (Fig.
In some of the specimens prepared with an Er:YAG laser (group 1 and group 2), a more uneven surface and deeper areas resembling craters were found, which were filled with the adhesive system (Fig.
By using digital magnification in the CLSM software, we were able to perform a more detailed study of the micromorphology of the fluorescently labeled objects, which was particularly useful in the control group. Fig.
CLSM image of a sample (group 1). Images of channel 1 (primer in red), channel 2 (bond in green) and superimposed image between them (in yellow) (×10).
CLSM image of a sample of group 2. Images of channel 1 (primer in red), channel 2 (bond in green) and superimposed image between them (in yellow) (×10).
CLSM image of a sample from group 3 (controls). Images of channel 1 (primer in red), channel 2 (bond in green) and superimposed image between them (×10).
“Craters”. CLSM image of a sample from group 1. Images of channel 1 (primer in red), channel 2 (bond in green) and superimposed image between them (×10).
Several studies have examined the specifics of working with Er:YAG lasers both in terms of dentin and enamel.[
The craters visualized in our study are also found by Yamada et al. in a preparation with an Er:YAG laser in the dentin without the presence of carbonization of the collagen structure.[
The micromorphological structure observed with CLSM in our study in the samples from groups 1 and 2, showing increased roughness and the presence of craters and long resin tags, could be considered as a favorable factor for increasing the strength of the adhesive bond, despite the report of some studies of other authors. According to Ceballo et al., Er:YAG lasers create a laser-modified layer that adversely affects dentin adhesion, so it is not an alternative bonding strategy to conventional acid etching. Acid etching of dentin, according to them, gives values of shear bond strength, which are significantly higher than those achieved only by laser ablation or laser ablation in combination with acid etching.[
We noticed extremely thick resin-tags at the bottom of the “crater” (Fig.
Based on our morphological study and the analysis of the dentinal surface performed with CLSM, we conclude that regarding dentin infiltration and hybrid layer formation on teeth prepared with Er:YAG laser, dentin has surfaces that are rougher than those prepared with metal burs. Shorter resin-tags were observed in the control group. When using Er:YAG lasers for preparation, crater-like irregularities of the surface were observed. As the application time of the self-etching primer increased twice, we could not detect a noticeable morphological change in the formed hybrid layer compared to the samples in which we applied an adhesive system according to the manufacturer’s instructions. The clinical role of the length and quality of tags as well as the ratio of infiltration must be further investigated.
This study was funded by the Medical University of Plov-div – Doctoral and postdoctoral projects 2020-2021; DPDP:12/2020.