Original Article |
Corresponding author: Desislava R. Makakova ( d.makakova@abv.bg ) © 2023 Desislava R. Makakova, Svetlin Alexandrov, Angelina Vlahova.
This is an open access article distributed under the terms of the Creative Commons Attribution License (CC BY 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Citation:
Makakova DR, Alexandrov S, Vlahova A (2023) Comparative analysis of tensile strength between three types of retraction cords. Folia Medica 65(3): 447-452. https://doi.org/10.3897/folmed.65.e83093
|
Introduction: Gingival retraction is the withdrawal of the marginal gingiva away from the tooth. This procedure creates a space between the prepared tooth and the gingival tissues to catch more fine details of the impression material. The most common retraction device used in clinical practice is the retraction cord.
Aim: The study aims to evaluate the tensile strength of retraction cords designed with different braiding technology.
Materials and methods: A total of 150 experimental units were studied. They were divided into 3 groups of 50 each according to the type of retraction cord (Ultrapak #00, braided cord without core, and braided cord with monofilament core). We tested the tensile strength in the LMT 100 micro-tensile apparatus. The data were analyzed using SPSS v. 21. A critical significance level of p<0.05 is used.
Results: The comparative analysis of the tensile strength of the retraction cords shows a statistically significant difference between the braided and the knitted threads (Ultrapak #00) (p<0.001). The results indicated the highest tensile strength of the cord with monofilament 41.95 N/mm2, followed closely by the cord with the same cotton braid without monofilament with average strength of 39.80 N/mm2, and last came the Ultrapak cord 22.11 N/mm2.
Conclusions: The braided retraction cords show higher tensile strength compared to the Ultrapak #00 cord, which is made using knitting technology.
gingival retraction, physical properties, retraction agents
The steps of making a fixed prosthesis include retraction of the gingival tissue, imprinting of the prosthetic field – prepared tooth, part or all of the row of teeth, together with the adjacent tissues, and registration of the antagonist teeth.[
The success of fixed prosthetic restorations depends largely on the health and stability of the surrounding periodontal structures.[
The use of materials with insufficient properties, as well as inappropriate technique for the retraction of the gingival pocket can lead to irreversible changes in the gingival tissues.[
The objective of this study was to evaluate the tensile strength of the two types of retraction cords made with different braiding technology and to compare it with a retraction cord from the market.
The retraction cords used in this study were Ultrapak #00 (U) – Ultradent Products, South Jordan, Utah, USA (Fig.
1. 100% cotton braided cords without core (BC)
2. 92% cotton and 8% polyamide braided cords with monofilament core (BCM).
The braided cords are produced on a circular braiding machine by Herzog (Germany) with 8 carriers using braiding system. The even number of braiding carriers shapes a round braid (Fig.
The item is made in raw form and is subjected to additional physicochemical treatment and sterilization. A total of 150 experimental units were made, 50 from each group of retraction cords – U, BC, and BCM, according to the silicone matrix we made (Silibest, BMS Italy). The prepared prototype units were fixed on a round polypropylene pad. A cuff of pink plaque wax 20 mm high and 1.5 cm thick was placed around the pad. The prepared setup was filled with laboratory silicone for model duplication (Elite Double 22, Zhermack Italy) (Fig.
After elasticizing the silicone mass, the wax prototypes were removed, and thus the finished shape for making the future prototypes was obtained. The tested retraction cords were placed in the vacated nests, and they were fixed with epoxy resin (Epovit, Vector) (Fig.
The tensile strength test was performed in a LMT 100 micro-tension apparatus (LAM Industry, Italy) (Fig.
The data were analyzed using SPSS v. 21. The statistical methods used to analyze the information are descriptive analysis by means of two-dimensional frequency distribution tables (cross-tabulation) and analysis of variance (one-way ANOVA). A critical significance level of 0.05 is used.
The comparison of the average strength of the three types of retraction cords shows that the cord with monofilament had the highest tensile strength of 41.5±0.47 N/mm2, followed closely by the cord with the same cotton braid without monofilament with average strength of 39.80±0.89 N/mm2, and the Ultrapak cord with strength of 22.11±0.62 N/mm2.
The average tensile strength of the experimental units was the highest in the monofilament cords compared to the other two types – 41.95±0.41 N/mm2. With 95% confidence interval, the average force required to break the connection of a monofilament-core cord was 41.11 N/mm2 as lower limit and 42.74 N/mm2 as upper limit. The minimum reported breaking force was 33.1 N/mm2 and the maximum 45.4 N/mm2 (Fig.
The average tensile strength of the experimental cord without a core was 39.80±0.89 N/mm2. With 95% confidence interval, the average force required to break the connection in the case of cotton thread without core was 38.04 N/mm2 as lower limit, and 41.57 N/mm2 as upper limit. The minimum reported breaking force was 23.6 N/mm2 and the maximum 50.8 N/mm2.
The average tensile strength of the experimental Ultrapak cords was the lowest (22.11±0.62 N/mm2) compared to the other two types. With a 95% confidence interval, the average force required to break the connection of the Ultrapak cord was 20.86 N/mm2 as lower limit and 23.36 N/mm2 as upper limit. The minimum reported breaking force was 13.0 N/mm2 and the maximum 30.5 N/mm2.
The multiple comparisons show a statistical difference between the compared groups according to numerical criterion. Distribution of the three groups, by analysis of variance (one-way ANOVA), confirmed a statistical difference in the group BCM and BC, where p=0.025. When comparing BCM and U, a statistically significant difference was found, where p<0.001, at F=267.92. We can also conclude that the distribution, by analysis of variance (one-way ANOVA), and in the comparison of BC and U, there was a statistical difference, where p<0.001, at F=267.92. The analysis revealed a statistically significant difference in the comparison of the tested test specimens of the three types of retraction cords (BC, BCM and U) for tensile strength.
Retraction cords are useful clinical aids, but there is insufficient information on the desired physical characteristics that would preserve their integrity during tension. This study aims to establish, under experimental conditions, the extent to which tensile strength is influenced by the braiding technology of the cords.
The obtained results show that the possible reasons for the higher values of the tensile strength of the two threads proposed by us (BS and BCM) compared to the Ultrapak cord can be found in their structure and composition. The applied effectiveness of the above retraction cords will have its real evidence after testing the two braided cords in a clinical environment.
One of the first studies to examine the physicochemical characteristics of retraction cords in more detail was by Nietro-Martinez et al. Their study established the effect of the impregnation substances and their concentrations on the tensile strength of the retraction cords, as well as the influence of the different diameters and number of threads of the cords. They come to the conclusion that the standard cotton cords are weakened when impregnated with aluminum sulfate and that they are even more weakened when using ferrous sulfate.[
The findings of another study by Madhok et al. are consistent with those of Nietro-Martinez et al.[
In another study by Jokstad among dentists and dental students comparing braided and knitted cords, the knitted retraction cord was better ranked than the braided, as one of the evaluation criteria in this study was whether the cord frays during placement.[
Gingival retraction is a routine procedure. However, it can cause gum injury due to improper application of the retraction techniques.[
The retraction cord must be strong enough to withstand the force of manipulating the gingival tissue, inserting and removing it from the gingival sulcus, in order to withdraw the gums before taking an impression.
Based on the results of the study, we come to the conclusion that the braided retraction cord with a monofilament core shows the highest tensile strength, followed by the braided thread without monofilament, while the knitted Ultrapak #00 has the lowest tensile strength. The difference is due to the structure and composition of the cords, which affect the durability of the retraction cords. Our results should be further tested in other non-clinical and clinical studies.
The authors have no support to report.
The authors have no funding to report.
The authors have declared that no competing interests exist.